Back to Search
Start Over
Infectious consequences of hematoma from cardiac implantable electronic device procedures and the role of the antibiotic envelope: A WRAP-IT trial analysis.
- Source :
- Heart Rhythm; Dec2021, Vol. 18 Issue 12, p2080-2086, 7p
- Publication Year :
- 2021
-
Abstract
- <bold>Background: </bold>Hematoma is a complication of cardiac implantable electronic device (CIED) procedures and may lead to device infection. The TYRX antibacterial envelope reduced major CIED infection by 40% in the randomized WRAP-IT (World-wide Randomized Antibiotic Envelope Infection Prevention Trial) study, but its effectiveness in the presence of hematoma is not well understood.<bold>Objective: </bold>The purpose of this study was to evaluate the incidence and infectious consequences of hematoma and the association between envelope use, hematomas, and major CIED infection among WRAP-IT patients.<bold>Methods: </bold>All 6800 study patients were included in this analysis (control 3429; envelope 3371). Hematomas occurring within 30 days postprocedure (acute) were characterized and grouped by study treatment and evaluated for subsequent infection risk. Data were analyzed using Cox proportional hazard regression modeling.<bold>Results: </bold>Acute hematoma incidence was 2.2% at 30 days, with no significant difference between treatment groups (envelope vs control hazard ratio [HR] 1.15; 95% confidence interval [CI] 0.84-1.58; P = .39). Through all follow-up, the risk of major infection was significantly higher among control patients with hematoma vs those without (13.1% vs 1.6%; HR 11.3; 95% CI 5.5-23.2; P <.001). The risk of major infection was significantly lower in the envelope vs control patients with hematoma (2.5% vs 13.1%; HR 0.18; 95% CI 0.04-0.85; P = .03).<bold>Conclusion: </bold>The risk of hematoma was 2.2% among WRAP-IT patients. Among control patients, hematoma carried a >11-fold risk of developing a major CIED infection. This risk was significantly mitigated with antibacterial envelope use, with an 82% reduction in major CIED infection among envelope patients who developed hematoma compared to control. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 15475271
- Volume :
- 18
- Issue :
- 12
- Database :
- Supplemental Index
- Journal :
- Heart Rhythm
- Publication Type :
- Academic Journal
- Accession number :
- 153495506
- Full Text :
- https://doi.org/10.1016/j.hrthm.2021.07.011