iPhone App compared with standard blood pressure measurement -The iPARR trial.

Authors :
Dörr, Marcus
Weber, Stefan
Birkemeyer, Ralf
Leonardi, Licia
Winterhalder, Clemens
Raichle, Christina J.
Brasier, Noé
Burkard, Thilo
Eckstein, Jens
Source :
American Heart Journal; Mar2021, Vol. 233, p102-108, 7p
Publication Year :
2021

Abstract

<bold>Background: </bold>The possibility to use built-in smartphone-cameras for photoplethysmographic (PPG) recording of pulse waves lead to the release of numerous health apps, claiming to measure blood pressure (BP) based on PPG signals. Even though these apps are highly popular, not a single one is clinically validated. Aim of the current study was to test systolic BP (sBP) estimation by a promising new algorithm in a large clinical setting.<bold>Methods: </bold>The study was designed based on the European Society of Hypertension International Protocol Revision 2010. Each individual received 7 sequential BP measurements, starting with the reference device - an automated oscillometric cuff device - followed by the PPG recording at the patients' index finger.<bold>Results: </bold>A total 1,036 subjects were recruited of which 965 could be included for final analysis leading to 2,895 pairs of comparison. Mean (±SD) error between test and reference device was -0.41 (±16.52) mmHg. Only 38.1% of all 2,895 BP comparisons reached a delta within ±5 mmHg, while 29.3% reached a delta larger than 15 mmHg. Bland-Altman plot showed an overestimation of smartphone sBP in comparison to reference sBP in low range and an underestimation in high sBP range.<bold>Conclusions: </bold>According to the European Society of Hypertension International Protocol Revision 2010 specifications the algorithm failed validation criteria for sBP measurement and was not commercialized. These findings emphasize that health apps should be rigorously validated according to common guidelines before market release as under- and/or overestimation of BP is potentially exposing persons at health risks in short and long term.<bold>Trial Registration: </bold>ClinicalTrials.gov, number NCT02552030. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00028703
Volume :
233
Database :
Supplemental Index
Journal :
American Heart Journal
Publication Type :
Academic Journal
Accession number :
148773199
Full Text :
https://doi.org/10.1016/j.ahj.2020.12.003