Open fracture of deep brain stimulation leads with normal electrical impedances.
- Source :
- Brain Stimulation; Nov2020, Vol. 13 Issue 6, p1639-1641, 3p
- Publication Year :
- 2020
-
Abstract
- • Normal-range electrical impedances can be measured from fractured DBS electrodes. • Only combining impedance measurements and imaging can rule out electrode fractures. • This is probably the first reported open DBS electrode fracture resulting from a fall. [ABSTRACT FROM AUTHOR]