Back to Search Start Over

Midterm Results of Delta Ceramic-on-Ceramic Total Hip Arthroplasty.

Authors :
Hamilton, William G.
McAuley, James P.
Blumenfeld, Thomas J.
Lesko, James P.
Himden, Sam E.
Dennis, Douglas A.
Source :
Journal of Arthroplasty; Sep2015 Supplement, Vol. 30 Issue 9, p110-115, 6p
Publication Year :
2015

Abstract

This study reports mid-term results of Delta ceramic on ceramic (COC) in total hip arthroplasty (THA). Subjects received Delta COC THA in a prospective multi-center study with either 28 mm ( n = 177) or 36 mm ( n = 168) articulations. Annual clinical and radiographic evaluations were performed, and patients were asked about hip noises. At mean 5.3-year follow-up there were 3 (0.9%) post-operative liner fractures. Nine revisions were performed (2 liner fracture, 4 stem loosening, 3 deep infection). Kaplan-Meier survivorship at 6 years was 96.9% (94.0–98.4). Twenty-six (7.5%) subjects reported squeaking, of whom none were revised. One (0.3%) subject could reproduce a sound in clinic. More patients reported squeaking with a 36 mm bearing (28 mm: 7/177, 36 mm: 19/168, P = 0.013). [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
08835403
Volume :
30
Issue :
9
Database :
Supplemental Index
Journal :
Journal of Arthroplasty
Publication Type :
Academic Journal
Accession number :
109278511
Full Text :
https://doi.org/10.1016/j.arth.2015.02.047