Back to Search
Start Over
Multi-time-lag PIV analysis of steady and pulsatile flows in a sidewall aneurysm.
- Source :
- Experiments in Fluids; Jun2014, Vol. 55 Issue 6, p1-11, 11p
- Publication Year :
- 2014
-
Abstract
- The effect of inflow waveform on the hemodynamics of a real-size idealized sidewall intracranial aneurysm (IA) model was investigated using particle imaging velocimetry (PIV). For this purpose, we implemented an error analysis based on several PIV measurements with different time lags to ensure high precision of velocity fields measured in both the IA and the parent artery. The relative error measured in the main part of the circulating volume was <1 % despite the three orders of magnitude difference of parent artery and IA dome velocities. Moreover, important features involved in IA evolution were potentially emphasized from the qualitative and quantitative flow pattern comparison resulting from steady and unsteady inflows. In particular, the flow transfer in IA and the vortical structure were significantly modified when increasing the number of harmonics for a typical physiological flow, in comparison with quasi-steady conditions. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 07234864
- Volume :
- 55
- Issue :
- 6
- Database :
- Complementary Index
- Journal :
- Experiments in Fluids
- Publication Type :
- Academic Journal
- Accession number :
- 96693754
- Full Text :
- https://doi.org/10.1007/s00348-014-1746-0