Back to Search
Start Over
Switched Bias Differential MOSFET Dosimeter.
- Source :
- IEEE Transactions on Nuclear Science; 6/10/2014 Part 2, Vol. 61 Issue 3, p1407-1413, 7p
- Publication Year :
- 2014
-
Abstract
- This paper presents a differential MOSFET sensor reading technique based on the bias controlled cycled measurement. The circuit was implemented, and tested with gamma radiation from a 60-cobalt source. Temperature rejection performance was assessed during the exposure in real-time measurements. The results show that in comparison with a single MOSFET dosimeter the thermal drift is 20 times smaller and the radiation sensitivity is approximately 10% higher. The switched biasing allows to extend the measurement range beyond MOSFET’s threshold voltage saturation. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00189499
- Volume :
- 61
- Issue :
- 3
- Database :
- Complementary Index
- Journal :
- IEEE Transactions on Nuclear Science
- Publication Type :
- Academic Journal
- Accession number :
- 96647340
- Full Text :
- https://doi.org/10.1109/TNS.2014.2316337