Analysis of signal coherence in ultrasound beamforming.

Authors :
Shougang Wang
Sheng-Wen Huang
Robert, Jean-Luc
Korukonda, Sanghamithra
Vignon, Francois
Erkamp, Ramon
Radulescu, Emil
Source :
2013 IEEE International Ultrasonics Symposium (IUS); 2013, p809-812, 4p
Publication Year :
2013

Abstract

Coherence factor weighting is used to suppress sidelobes, reduce artifacts, and improve the contrast and spatial resolution. The method provides a less hardware demanding solution with acceptable results. We conduct an analysis on coherence factor (CF) and a two-aperture coherence (TAC) estimation for the purpose of general understanding and fast estimation. A simulation was implemented based on linear description of ultrasound system and a simplified lateral scatterer distribution at the region of interest to analyze differences. Characters of CF and TAC for a point, homogeneous speckle, and a point and a cyst within homogeneous speckle environment under narrow and broad band were simulated. [ABSTRACT FROM PUBLISHER]