Back to Search
Start Over
A differential high-resolution motorized multi-projection approach for an experimental Magnetic Induction Tomography prototype.
- Source :
- 2012 IEEE International Symposium on Medical Measurements & Applications Proceedings; 1/ 1/2012, p1-4, 4p
- Publication Year :
- 2012
-
Abstract
- In this paper, the first experimental results for a multi-projection differential coil Magnetic Induction Tomography system are presented. All current systems are static, allowing only the acquisition of a small number of fixed projections. The presented setup provides a much higher density of projection measurements which reduces the ill-posedness of the inverse problem. This is further enhanced through the use of a topview camera to detect the external frontiers of the object. The experimental setup is based on a previously presented design and consists of a geometrically centered excitation and four pairs of anti-series differential coils equally spaced at 45°. Simulation and experimental results for 3D phantom structures are shown. [ABSTRACT FROM PUBLISHER]
Details
- Language :
- English
- ISBNs :
- 9781467308809
- Database :
- Complementary Index
- Journal :
- 2012 IEEE International Symposium on Medical Measurements & Applications Proceedings
- Publication Type :
- Conference
- Accession number :
- 86567941
- Full Text :
- https://doi.org/10.1109/MeMeA.2012.6226664