Back to Search Start Over

First measurement of neutron emission profile on JT-60U using Stilbene neutron detector with neutron-gamma discrimination.

Authors :
Ishikawa, M.
Nishitani, T.
Morioka, A.
Takechi, M.
Shinohara, K.
Shimada, M.
Miura, Y.
Nagami, M.
Kaschuck, Yu. A.
Source :
Review of Scientific Instruments; Dec2002, Vol. 73 Issue 12, p4237, 6p
Publication Year :
2002

Abstract

This article describes a diagnostic for measuring neutron emission profile in JT-60U. The Stilbene neutron detector, developed by TRINITI laboratory in Russia, has been installed on the JT-60U Tokamak to measure the neutron emission profile for the first time. The Stilbene neutron detector is a detector which combines a Stilbene crystal scintillator with a neutron-gamma pulse shape discrimination circuit, with a very compact size. Performance tests were carried out using neutron and gamma-ray source prior to installation on JT-60U. Good gamma suppression of the Stilbene neutron detector was verified. Though the neutron emission profile obtained by Stilbene neutron detectors has error of 30% in innermost channel with a calculation using measured plasma parameters, there is an agreement within 10% error in the other channels. [ABSTRACT FROM AUTHOR]

Subjects

Subjects :
NEUTRON counters
NEUTRONS
TOKAMAKS

Details

Language :
English
ISSN :
00346748
Volume :
73
Issue :
12
Database :
Complementary Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
8563473