First measurement of neutron emission profile on JT-60U using Stilbene neutron detector with neutron-gamma discrimination.
- Source :
- Review of Scientific Instruments; Dec2002, Vol. 73 Issue 12, p4237, 6p
- Publication Year :
- 2002
-
Abstract
- This article describes a diagnostic for measuring neutron emission profile in JT-60U. The Stilbene neutron detector, developed by TRINITI laboratory in Russia, has been installed on the JT-60U Tokamak to measure the neutron emission profile for the first time. The Stilbene neutron detector is a detector which combines a Stilbene crystal scintillator with a neutron-gamma pulse shape discrimination circuit, with a very compact size. Performance tests were carried out using neutron and gamma-ray source prior to installation on JT-60U. Good gamma suppression of the Stilbene neutron detector was verified. Though the neutron emission profile obtained by Stilbene neutron detectors has error of 30% in innermost channel with a calculation using measured plasma parameters, there is an agreement within 10% error in the other channels. [ABSTRACT FROM AUTHOR]
- Subjects :
- NEUTRON counters
NEUTRONS
TOKAMAKS
Subjects
Details
- Language :
- English
- ISSN :
- 00346748
- Volume :
- 73
- Issue :
- 12
- Database :
- Complementary Index
- Journal :
- Review of Scientific Instruments
- Publication Type :
- Academic Journal
- Accession number :
- 8563473
- Full Text :
- https://doi.org/10.1063/1.1518145