Electrical-thermal performance of a cooled RF applicator for hepatic ablation with additional distant infusion of hypertonic saline: In vivo study and preliminary computer modeling.

Authors :
Romero-Méndez, Ricardo
Tobajas, Pilar
Burdío, Fernando
Gonzalez, Ana
Navarro, Ana
Grande, Luis
Berjano, Enrique
Source :
International Journal of Hyperthermia; Nov2012, Vol. 28 Issue 7, p653-662, 10p
Publication Year :
2012

Abstract

Purpose: The Cool-tip electrode is one of the most widely employed applicators in radiofrequency (RF) hepatic ablation. Previous research demonstrated that it is possible to enlarge coagulation volume when the single cooled electrode is associated with distant infusion of saline (hybrid applicator). The aim of this study was to compare the electrical-thermal behaviour of the Cool-tip electrode with that of the hybrid applicator. Materials and methods: Forty-two RF ablations were performed on a total of 10 pigs: 22 with the Cool-tip electrode and 20 with the hybrid applicator (low infused saline volumetric flow rate of 6 mL/h at 2 mm distance). We compared both electrical performance (delivered power and number of roll-offs, i.e. sudden rises in impedance that interrupt the power delivery) and coagulation zone characteristics. In addition, we built a one-dimensional model to provide a basic physical explanation of the difference in performance between the different applicators. Results: The experimental results showed that the number of roll-offs with the Cool-tip electrode was higher (24.3 ± 3.1 versus 6.7 ± 7.0). The hybrid applicator created larger coagulation volumes (19.7 ± 9.5 cm<superscript>3</superscript> versus 9.5 ± 5.8 cm<superscript>3</superscript>) with larger transverse diameters (2.5 ± 0.6 versus 1.9 ± 0.5 cm). The one-dimensional model confirmed the delay in the incidence of the first roll-off, but not the heterogeneity of the hybrid applicator's electrical performance in the experiments. Conclusions: The hybrid applicator produces fewer roll-off episodes than the Cool-tip electrode and creates larger coagulation volumes with larger transverse diameters. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02656736
Volume :
28
Issue :
7
Database :
Complementary Index
Journal :
International Journal of Hyperthermia
Publication Type :
Academic Journal
Accession number :
82468866
Full Text :
https://doi.org/10.3109/02656736.2012.711894