Evaluating phantom image quality parameters to optimise patient radiation dose in dental digital radiology.

Authors :
Gonzalez, Luciano
Vano, Eliseo
Fernandez, Ramon
Ziraldo, Viviana
Delgado, Jaime
Delgado, Victor
Moro, Javier
Ubeda, Carlos
Source :
Radiation Protection Dosimetry; Aug2012, Vol. 151 Issue 1, p95-101, 7p
Publication Year :
2012

Abstract

Our objective was to obtain images of a predictable level of quality using an intraoral X-ray system with digital imaging, avoiding patient overexposure. A polymethylmethacrylate (PMMA) physical test phantom was imaged at different exposure times and at various PMMA thicknesses using a dental imaging coupled charge device. Two identical regions of interest (ROIs) were chosen in every image file, and quality was numerically evaluated by measuring high-contrast spatial resolutions, low-contrast thresholds and signal-to-noise ratios. In addition, three practitioners proposed personal quality scores by image inspection. Numerical contents in the ROIs, related to the image quality, were plotted against exposure time. From here, a simple expression linking the exposure time with the thickness to obtain images of comparable quality was deduced. As a result, the optimum exposure time for imaging with a predictable level of quality can be inferred. The potential effect could imply savings above 1000 man Sv, roughly 20 % of the collective dose due to dental imaging, over a population of 1540 millions. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01448420
Volume :
151
Issue :
1
Database :
Complementary Index
Journal :
Radiation Protection Dosimetry
Publication Type :
Academic Journal
Accession number :
78359714
Full Text :
https://doi.org/10.1093/rpd/ncr470