Some applications of enthalpimetric detection in flow injection analysis.

Authors :
Linden, W.
Bos, M.
Heskamp, H.
Wilms, H.
Source :
Fresenius' Zeitschrift für Analytische Chemie; 1987, Vol. 329 Issue 4, p440-443, 4p
Publication Year :
1987

Abstract

A flow injection manifold is described based on a flow-through arrangement of thermistors for measuring heat effects of chemical reactions. Temperature changes down to 10 °C can be measured reproducibly. The working range for the determination of a component is dependent on the reaction enthalpy and can vary from 10 to 10 mol/l for acids and bases, from 10 to 1000 ppm for e.g., alkylhydroperoxides. Sample frequencies up to 150 samples per hour can be realized. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00161152
Volume :
329
Issue :
4
Database :
Complementary Index
Journal :
Fresenius' Zeitschrift für Analytische Chemie
Publication Type :
Academic Journal
Accession number :
72809731
Full Text :
https://doi.org/10.1007/BF00480081