Assessment of thrombocytopenic disorders using the Platelet Function Analyzer (PFA-100® )*.
- Source :
- British Journal of Haematology; Jun2002, Vol. 117 Issue 4, p961-964, 4p, 6 Graphs
- Publication Year :
- 2002
-
Abstract
- Summary. The Platelet Function Analyzer (PFA-100<superscript>®</superscript> ) was used to measure platelet function in paediatric patients with destructive versus underproduction thrombocytopenia. Closure time (CT) and total volume (TV) measurements with standard 150 μm apertures discriminated between patients with similar platelet counts from 30 to 150 × 10<superscript>9</superscript> /l. However, at platelet counts < 30 × 10<superscript>9</superscript> /l, a 100-μm aperture (experimental) gave the best assessment of platelet function. TV results could be analysed even when CTs were indeterminate. Further investigations are warranted to more fully understand the relationships among platelet function as measured by the PFA-100<superscript>®</superscript> in standard/experimental modes, bleeding and transfusion outcome in thrombocytopenia. [ABSTRACT FROM AUTHOR]
- Subjects :
- BLOOD platelets
THROMBOCYTOPENIA
Subjects
Details
- Language :
- English
- ISSN :
- 00071048
- Volume :
- 117
- Issue :
- 4
- Database :
- Complementary Index
- Journal :
- British Journal of Haematology
- Publication Type :
- Academic Journal
- Accession number :
- 6808173
- Full Text :
- https://doi.org/10.1046/j.1365-2141.2002.03511.x