Back to Search Start Over

Design and tests of high sensitivity NTD Ge thermometers for the Planck-High Frequency Instrument.

Authors :
Piat, M.
Torre, J. P.
Lamarre, J. M.
Beeman, J. W.
Bhatia, R. S.
Leriche, B.
Crussaire, J. P.
Langlet, F.
Source :
AIP Conference Proceedings; 2002, Vol. 605 Issue 1, p79, 4p
Publication Year :
2002

Abstract

The High Frequency Instrument of Planck needs high sensitivity semi-conductors at low temperature to monitor the temperature of the bolometer plate. We have modeled such thermometers by using a semi-analytical approach of Anderson insulators, taking into account both the electrical field and the electron/phonon decoupling effects. The optimized design uses convenient NTD Ge material and has larger dimension than the initial design. The first measurements of these optimized thermometers showed a significant thermal de-coupling effect due to Kapitza resistance with its mechanical support. Nevertheless, a sensitivity of about 8nK.Hz[sup -0.5], not far from the predicted one, was obtained. The noise spectrum of the thermometer was flat down to 1Hz, dominated at lower frequency by the thermal fluctuations. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0094243X
Volume :
605
Issue :
1
Database :
Complementary Index
Journal :
AIP Conference Proceedings
Publication Type :
Conference
Accession number :
6310308