Back to Search
Start Over
Design and tests of high sensitivity NTD Ge thermometers for the Planck-High Frequency Instrument.
- Source :
- AIP Conference Proceedings; 2002, Vol. 605 Issue 1, p79, 4p
- Publication Year :
- 2002
-
Abstract
- The High Frequency Instrument of Planck needs high sensitivity semi-conductors at low temperature to monitor the temperature of the bolometer plate. We have modeled such thermometers by using a semi-analytical approach of Anderson insulators, taking into account both the electrical field and the electron/phonon decoupling effects. The optimized design uses convenient NTD Ge material and has larger dimension than the initial design. The first measurements of these optimized thermometers showed a significant thermal de-coupling effect due to Kapitza resistance with its mechanical support. Nevertheless, a sensitivity of about 8nK.Hz[sup -0.5], not far from the predicted one, was obtained. The noise spectrum of the thermometer was flat down to 1Hz, dominated at lower frequency by the thermal fluctuations. [ABSTRACT FROM AUTHOR]
- Subjects :
- THERMOMETERS
SEMICONDUCTORS
MATERIALS at low temperatures
Subjects
Details
- Language :
- English
- ISSN :
- 0094243X
- Volume :
- 605
- Issue :
- 1
- Database :
- Complementary Index
- Journal :
- AIP Conference Proceedings
- Publication Type :
- Conference
- Accession number :
- 6310308