Clinical value, normative retinal sensitivity values, and intrasession repeatability using a combined spectral domain optical coherence tomography/scanning laser ophthalmoscope microperimeter.

Authors :
Anastasakis, A.
McAnany, J. J.
Fishman, G. A.
Seiple, W. H.
Source :
Eye; Feb2011, Vol. 25 Issue 2, p245-251, 7p
Publication Year :
2011

Abstract

PurposeTo establish normative values for macular light sensitivity and to determine the intrasession fluctuation of perimetric responses using the OPKO/OTI microperimeter.MethodsA total of 32 visually normal subjects participated in the study. A standardized grid pattern was used for testing, which consisted of 28 points arranged concentrically in three circles that occupied an area of 11° (in diameter) within the central macula. Each subject participated in at least two tests. Parameters evaluated included: overall mean macular sensitivity for test 1 and 2, overall difference in mean macular sensitivity between tests, and the mean sensitivity for each circle. The relationship between sensitivity and age was also examined.ResultsThe overall median sensitivity for test 1 was 16.8 decibels (dB) and for test 2 was 16.9 dB. The median sensitivities for test 1 and test 2 were not significantly different (P=0.72). The mean intrasession sensitivity difference was 0.13 dB. The variability of the sensitivity difference between tests decreased as mean sensitivity increased. The sensitivity values averaged across the two tests for inner, middle, and outer circles ranged from 14.3 to 18.8 dB (median value of 16.9 dB), 13.8-18.3 dB (median value of 17.2 dB), and 11.3-18.3 dB (median value of 16.6 dB), respectively. Linear regression analysis showed a 0.5 dB sensitivity loss for each decade of life.ConclusionWe documented a narrow range of intrasession fluctuation using the OPKO/OTI microperimeter. The establishment of normative sensitivity values will facilitate monitoring the loss of macular visual function in patients with retinal disease. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0950222X
Volume :
25
Issue :
2
Database :
Complementary Index
Journal :
Eye
Publication Type :
Academic Journal
Accession number :
57985727
Full Text :
https://doi.org/10.1038/eye.2010.158