Back to Search Start Over

First certified reference material for cystatin C in human serum ERM-DA471/IFCC.

Authors :
Grubb, Anders
Blirup-Jensen, Søren
Lindström, Veronica
Schmidt, Camilla
Althaus, Harald
Zegers, Ingrid
Source :
Clinical Chemistry & Laboratory Medicine; Nov2010, Vol. 48 Issue 11, p1619-1621, 3p
Publication Year :
2010

Abstract

The IFCC Working Group for the Standardisation of Cystatin C (WG-SCC), in collaboration with the Institute for Reference Materials and Measurements (IRMM), announces the availability of the new certified reference material ERM-DA471/IFCC. The material was characterised using a pure protein primary reference preparation (PRP) as calibrant. The PRP was prepared from recombinant cystatin C, and its concentration measured using dry mass determination. The characterisation of ERM-DA471/IFCC was performed by particle enhanced immuno-nephelometry, particle enhanced immuno-turbidimetry, and enzyme amplified single radial immuno-diffusion. The certified cystatin C mass concentration in ERM-DA471/IFCC, if reconstituted according to the specified procedure, is 5.48 mg/L, the expanded uncertainty ( k=2) being 0.15 mg/L. Clin Chem Lab Med 2010;48:1619-21. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
14346621
Volume :
48
Issue :
11
Database :
Complementary Index
Journal :
Clinical Chemistry & Laboratory Medicine
Publication Type :
Academic Journal
Accession number :
55090521
Full Text :
https://doi.org/10.1515/CCLM.2010.318