- The IFCC Working Group for the Standardisation of Cystatin C (WG-SCC), in collaboration with the Institute for Reference Materials and Measurements (IRMM), announces the availability of the new certified reference material ERM-DA471/IFCC. The material was characterised using a pure protein primary reference preparation (PRP) as calibrant. The PRP was prepared from recombinant cystatin C, and its concentration measured using dry mass determination. The characterisation of ERM-DA471/IFCC was performed by particle enhanced immuno-nephelometry, particle enhanced immuno-turbidimetry, and enzyme amplified single radial immuno-diffusion. The certified cystatin C mass concentration in ERM-DA471/IFCC, if reconstituted according to the specified procedure, is 5.48 mg/L, the expanded uncertainty ( k=2) being 0.15 mg/L. Clin Chem Lab Med 2010;48:1619-21. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 14346621
- Volume :
- 48
- Issue :
- 11
- Database :
- Complementary Index
- Journal :
- Clinical Chemistry & Laboratory Medicine
- Publication Type :
- Academic Journal
- Accession number :
- 55090521
- Full Text :
- https://doi.org/10.1515/CCLM.2010.318