Back to Search
Start Over
Determination of the selenium isotopic compositions in Se-rich yeast by hydride generation-inductively coupled plasma multicollector mass spectrometry.
- Source :
- JAAS (Journal of Analytical Atomic Spectrometry); Nov2010, Vol. 25 Issue 11, p1695-1703, 9p
- Publication Year :
- 2010
-
Abstract
- A method was developed for the precise determination of selenium isotopic compositions (δ82/77Se, δ82/76Se and δ82/74Se) by hydride generation inductively coupled plasma multiple collector mass spectrometry (HG-MC ICP MS) in Se-rich yeast. The conversion of all the selenium forms into Se(iv) during digestion and reduction steps was investigated by anion-exchange ICP quadrupole MS (SAX ICP Q MS) and was found to be complete. Measurement precision values expressed as deltas (δ) with regard to NIST SRM3149 were between 0.2–0.4‰ (2SD) over a period of 6 months for 25 or 50 ng mL−1of introduced selenium. The application of the method to the analysis of nine Se-rich yeast supplements from different manufactures showed significant variations in the isotope ratios suggesting that the latter can be an attractive parameter for the yeast provenance discrimination. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 02679477
- Volume :
- 25
- Issue :
- 11
- Database :
- Complementary Index
- Journal :
- JAAS (Journal of Analytical Atomic Spectrometry)
- Publication Type :
- Academic Journal
- Accession number :
- 54844616
- Full Text :
- https://doi.org/10.1039/c004933f