Back to Search Start Over

First results obtained from the soft x-ray pulse height analyzer on experimental advanced superconducting tokamak.

Authors :
Xu, P.
Lin, S. Y.
Hu, L. Q.
Duan, Y. M.
Zhang, J. Z.
Chen, K. Y.
Zhong, G. Q.
Source :
Review of Scientific Instruments; Jun2010, Vol. 81 Issue 6, p063501, 5p, 3 Diagrams, 2 Charts, 6 Graphs
Publication Year :
2010

Abstract

An assembly of soft x-ray pulse height analyzer system, based on silicon drift detector (SDD), has been successfully established on the experimental advanced superconducting tokamak (EAST) to measure the spectrum of soft x-ray emission (E=1–20 keV). The system, including one 15-channel SDD linear array, is installed on EAST horizontal port C. The time-resolved radial profiles of electron temperature and K<subscript>α</subscript> intensities of metallic impurities have been obtained with a spatial resolution of around 7 cm during a single discharge. It was found that the electron temperatures derived from the system are in good agreement with the values from Thomson scattering measurements. The system can also be applied to the measurement of the long pulse discharge for EAST. The diagnostic system is introduced and some typical experimental results obtained from the system are also presented. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
81
Issue :
6
Database :
Complementary Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
51848929
Full Text :
https://doi.org/10.1063/1.3443572