Accuracy of a Manual Torque Application Device for Morse-Taper Implants: A Technical Note.
- Source :
- International Journal of Oral & Maxillofacial Implants; Sep/Oct2004, Vol. 19 Issue 5, p743-748, 6p, 2 Color Photographs, 1 Chart, 2 Graphs
- Publication Year :
- 2004
-
Abstract
- Purpose: The objective of this study was to compare torques applied by new and used manual torque devices for Morse-taper implants. Material and Methods: Fifteen ITI manual torque devices were tested. Those in group 1 (n = 5) were new (ie, never used), those in group 2 (n = 5) had been used 50 to 200 times, and those in group 3 (n = 5) had been used 500 to 1,000 times. The torques applied by each device were measured for 35 Ncm and 15 Ncm targets in an experimental setup by a custom-made wrench with strain gauges connected to a data acquisition system. The strain-gauge signals were simultaneously delivered to a computer at a sample rate of 10,000 Hz and converted to torque units. Results: New devices applied higher torques than used devices for the 35-Ncm torque target (P <.05). The torques applied by group 3 devices were approximately 1.5 Ncm lower than those of other groups for the 35-Ncm target and approximately 1 Ncm lower for the 15-Ncm target. Discussion and Conclusion: ITI manual torque devices deliver consistent torque output, although a slight decrease occurs as a consequence of clinical use. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 08822786
- Volume :
- 19
- Issue :
- 5
- Database :
- Complementary Index
- Journal :
- International Journal of Oral & Maxillofacial Implants
- Publication Type :
- Academic Journal
- Accession number :
- 36846326