Developmental Validation of the Quantifiler® Duo DNA Quantification Kit for Simultaneous Quantification of Total Human and Human Male DNA and Detection of PCR Inhibitors in Biological Samples.
- Source :
- Journal of Forensic Sciences; Mar2009, Vol. 54 Issue 2, p305-319, 15p, 9 Charts, 11 Graphs
- Publication Year :
- 2009
-
Abstract
- The Quantifiler<superscript>®</superscript> Duo DNA Quantification kit enables simultaneous quantification of human DNA and human male DNA as well as detection of inhibitors of PCR in a single real-time PCR well. Pooled human male genomic DNA is used to generate standard curves for both human (ribonuclease P RNA component H1) and human male (sex determining region Y) specific targets. A shift in the cycle threshold ( C<subscript>T</subscript>) values for the internal positive control monitors the presence of PCR inhibitors in a sample. The assay is human specific and exhibits a high dynamic range from 0.023 to 50 ng/μL. In addition, the multiplex assay can detect as little as 25 pg/μL of human male DNA in the presence of a 1000-fold excess of human female DNA. The multiplex assay provides assessment of the DNA extract and guidance for the selection of the appropriate AmpFℓSTR<superscript>®</superscript> Amplification Kit to obtain interpretable short tandem repeat profiles. [ABSTRACT FROM AUTHOR]