Hand-Held Gamma-Ray Spectrometer Based on High-Efficiency Frisch-Ring CdZnTe Detectors.
- Source :
- IEEE Transactions on Nuclear Science; Oct2008 Part 2 of 2, Vol. 55, p2765-2769, 5p, 3 Black and White Photographs, 4 Graphs
- Publication Year :
- 2008
-
Abstract
- Frisch-ring CdZnTe detectors have demonstrated both good energy resolution, < 1% FWHM at 662 keV, and good efficiency in detecting gamma rays, highlighting the strong potential of CdZnTe materials for such applications. We are designing a hand-held gamma-ray spectrometer based on Frisch-ring detectors at Brookhaven National Laboratory. It employs an 8 × 8 CdZnTe detector array to achieve a high volume of 19.2 cm³, so greatly improving detection efficiency. By using the front-end application-specific integrated circuits (ASICs) developed at BNL, this spectrometer has a small profile and high energy-resolution. It includes a signal processing circuit, digitization and storage circuits, a high-voltage module, and a universal serial bus (USB) interface. In this paper, we detail the system's structure and report the results of our tests with it. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00189499
- Volume :
- 55
- Database :
- Complementary Index
- Journal :
- IEEE Transactions on Nuclear Science
- Publication Type :
- Academic Journal
- Accession number :
- 35946791
- Full Text :
- https://doi.org/10.1109/TNS.2008.2004803