Back to Search Start Over

Hand-Held Gamma-Ray Spectrometer Based on High-Efficiency Frisch-Ring CdZnTe Detectors.

Authors :
Cui, Yonggang
Bolotnikov, Aleksey
Camarda, Giuseppe
Hossain, Anwar
James, Ralph B.
De Geronimo, Gianluigi
Fried, Jack
O'Connor, Paul
Kargar, Alireza
Harrison, Mark J.
McGregor, Douglas S.
Source :
IEEE Transactions on Nuclear Science; Oct2008 Part 2 of 2, Vol. 55, p2765-2769, 5p, 3 Black and White Photographs, 4 Graphs
Publication Year :
2008

Abstract

Frisch-ring CdZnTe detectors have demonstrated both good energy resolution, < 1% FWHM at 662 keV, and good efficiency in detecting gamma rays, highlighting the strong potential of CdZnTe materials for such applications. We are designing a hand-held gamma-ray spectrometer based on Frisch-ring detectors at Brookhaven National Laboratory. It employs an 8 × 8 CdZnTe detector array to achieve a high volume of 19.2 cm³, so greatly improving detection efficiency. By using the front-end application-specific integrated circuits (ASICs) developed at BNL, this spectrometer has a small profile and high energy-resolution. It includes a signal processing circuit, digitization and storage circuits, a high-voltage module, and a universal serial bus (USB) interface. In this paper, we detail the system's structure and report the results of our tests with it. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189499
Volume :
55
Database :
Complementary Index
Journal :
IEEE Transactions on Nuclear Science
Publication Type :
Academic Journal
Accession number :
35946791
Full Text :
https://doi.org/10.1109/TNS.2008.2004803