Back to Search Start Over

Sensitive HPLC–ESI-MS Method for the Determination of Tiotropium in Human Plasma.

Authors :
Li Ding
Wenming Tan
Yindi Zhang
Jianping Shen
Zhengxing Zhang
Source :
Journal of Chromatographic Science; May2008, Vol. 46 Issue 5, p445-449, 5p
Publication Year :
2008

Abstract

A sensitive high-performance liquid chromatography–electrospray ionization-mass spectrometry (HPLC–ESI-MS) assay is established for the determination of tiotropium in human plasma using benzyltriethylammonium chloride as the internal standard (IS). After being treated with C18 cartridges, plasma samples are separated by HPLC on a reversed-phase C18 column with a mobile phase of 40mM ammonium acetate buffer–methanol (56:44, v/v). Tiotropium is determined in a single-quadrupole MS. HPLC–ESI-MS is performed in the selected ion monitoring mode using target ions at m/z 392.0 for tiotropium and m/z 192.3 for the IS. The calibration curve is linear over the range 1.5–30 pg/mL. The intra- and inter-assay variability values are less than 10.1% and 13.6%, respectively. The mean plasma extraction recovery of tiotropium is 92.3 ± 5.0%. The method has been successfully applied to studying the pharmacokinetics of tiotropium in healthy Chinese volunteers. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00219665
Volume :
46
Issue :
5
Database :
Complementary Index
Journal :
Journal of Chromatographic Science
Publication Type :
Academic Journal
Accession number :
32952798
Full Text :
https://doi.org/10.1093/chromsci/46.5.445