Determination of exhaled nitric oxide distributions in a diverse sample population using tunable diode laser absorption spectroscopy.
- Source :
- Applied Physics B: Lasers & Optics; Nov2006, Vol. 85 Issue 2/3, p427-435, 9p, 1 Color Photograph, 2 Diagrams, 1 Chart, 8 Graphs
- Publication Year :
- 2006
-
Abstract
- A liquid-nitrogen free mid-infrared tunable diode laser absorption spectroscopy (TDLAS) system equipped with a folded-optical-path astigmatic Herriott cell was used to measure levels of exhaled nitric oxide (eNO) and exhaled carbon dioxide (eCO<subscript>2</subscript>) in breath. Quantification of absolute eNO concentrations was performed using NO/CO<subscript>2</subscript> absorption ratios measured by the TDLAS system coupled with absolute eCO<subscript>2</subscript> concentrations measured with a non-dispersive infrared sensor. This technique eliminated the need for routine calibrations using standard cylinder gases. The TDLAS system was used to measure eNO in children and adults (n=799, ages 5 to 64) over a period of more than one year as part of a field study. Volunteers for the study self-reported data including age, height, weight, and health status. The resulting data were used to assess system performance and to generate eNO and eCO<subscript>2</subscript> distributions, which were found to be log-normal and Gaussian, respectively. There were statistically significant differences in mean eNO levels for males and females as well as for healthy and steroid naïve asthmatic volunteers not taking corticosteroid therapies. Ambient NO levels affected measured eNO concentrations only slightly, but this effect was not statistically significant. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 09462171
- Volume :
- 85
- Issue :
- 2/3
- Database :
- Complementary Index
- Journal :
- Applied Physics B: Lasers & Optics
- Publication Type :
- Academic Journal
- Accession number :
- 32776573
- Full Text :
- https://doi.org/10.1007/s00340-006-2301-3