Back to Search
Start Over
Measurement of derivative of ion temperature using high spatial resolution charge exchange spectroscopy with space modulation optics.
- Source :
- Review of Scientific Instruments; May2008, Vol. 79 Issue 5, p053506, 6p, 1 Diagram, 6 Graphs
- Publication Year :
- 2008
-
Abstract
- A new technique to measure the first and second derivatives of the ion temperature profile has been developed by using a charge exchange spectroscopy system with space modulation optics. The space observed is scanned up to ±3 cm with a cosine wave modulation frequency up to 30 Hz by shifting the object lens in front of the optical fiber bundle by 0.5 mm with a piezoelement. The first and second derivatives of ion temperature are derived from the modulation component of the ion temperature measured by using Fourier series expansion. [ABSTRACT FROM AUTHOR]
- Subjects :
- OPTICAL fibers
SPECTRUM analysis
IONS
TEMPERATURE measurements
CHARGE exchange
Subjects
Details
- Language :
- English
- ISSN :
- 00346748
- Volume :
- 79
- Issue :
- 5
- Database :
- Complementary Index
- Journal :
- Review of Scientific Instruments
- Publication Type :
- Academic Journal
- Accession number :
- 32548908
- Full Text :
- https://doi.org/10.1063/1.2921702