Rapid Communication Effects of Steris 1™Sterilization and Cidex®Ortho-Phthalaldehyde High-Level Disinfection on Durability of New-Generation Flexible Ureteroscopes.

Authors :
Jose Benito A. Abraham
Corollos S. Abdelshehid
Hak J. Lee
Geoffrey N. Box
Leslie A. Deane
Todd Le
Forrest Jellison
James F. Borin
Anthony Manipon
Elspeth M. McDougall
Ralph V. Clayman
Source :
Journal of Endourology; Sep2007, Vol. 21 Issue 9, p985-992, 8p
Publication Year :
2007

Abstract

Background and Purpose The effects of commonly used reprocessing methods on flexible ureteroscope longevity have never been examined. We prospectively studied the effects of Steris 1™sterilization and Cidex®ortho-phthalaldehyde (OPA) high-level disinfection (HLD) on the image quality, physical structure, and deflective properties of two new flexible ureteroscopes.Materials and Methods Two identical “out-of-the-box” Storz 11278AU1 flexible ureteroscopes (Karl Storz Endoscopy, Tuttlingen, Germany) were sterilized individually using the Steris 1™system (Steris Mentor, Ohio) or disinfected with Cidex®OPA (Advanced Sterilization Products, J&J, Irvine, CA) for 100 trials followed by a crossover to the other method for another 100 trials over a period of 1 year. After every five trials, optical quality, angle of deflection, and fiber damage were analyzed in the laboratory. Throughout the study, neither of these ureteroscopes was used clinically.Results After 100 trials, ureteroscope 1, which was sterilized initially in the Steris system, had a 12-mm tear on its shaft (noted after the 17thtrial), 297 damaged fibers, and a 37 drop in resolution (loss of 3.75 linesmm). There was no change in deflection from baseline. In contrast, after 100 cycles, ureteroscope 2, which was subjected to HLD with Cidex®OPA, had no visible external damage, a 0 change in resolution, 10 damaged fibers, and no change in deflection. After the crossover, ureteroscope 2 developed a semilunar defect that obscured the endoscopic view, whereas there was no further significant damage to ureteroscope 1.Conclusion After 100 cycles, the Steris 1 system rendered the flexible ureteroscope unusable, whereas HLD with Cidex®OPA had minimal adverse impact. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
08927790
Volume :
21
Issue :
9
Database :
Complementary Index
Journal :
Journal of Endourology
Publication Type :
Academic Journal
Accession number :
27099284
Full Text :
https://doi.org/10.1089/end.2007.0181