Back to Search Start Over

Cortical blood flow assessment with frequency-domain laser Doppler microscopy.

Authors :
Michael Atlan
Benoı⁁t C. Forget
Albert C. Boccara
Tania Vitalis
Armelle Rancillac
Andrew K. Dunn
Michel Gross
Source :
Journal of Biomedical Optics; Mar2007, Vol. 12 Issue 2, p24019-24019, 1p
Publication Year :
2007

Abstract

We report the assessment of cerebral blood flow (CBF) changes with a wide-field laser Doppler imager based on a CCD camera detection scheme, in vivo, in mice. The setup enables the acquisition of data in minimally invasive conditions. In contrast with conventional laser Doppler velocimeters and imagers, the Doppler signature of moving scatterers is measured in the frequency domain, by detuning a heterodyne optical detection. The quadratic mean of the measured frequency shift is used as an indicator of CBF. We observe a significant variability of this indicator in an experiment designed to induce blood flow changes. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
10833668
Volume :
12
Issue :
2
Database :
Complementary Index
Journal :
Journal of Biomedical Optics
Publication Type :
Academic Journal
Accession number :
26289675
Full Text :
https://doi.org/10.1117/1.2715184