Back to Search
Start Over
PATIENT DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY PROCEDURES IN KOREA.
- Source :
- Radiation Protection Dosimetry; 2007, Vol. 123 Issue 4, p540-545, 6p, 12 Charts
- Publication Year :
- 2007
-
Abstract
- This study is the first nationwide investigation aimed at estimating the patient dose for radiographic examinations in Korea including gastrointestinal studies, computed tomography and mammography. The survey data from 161 hospitals and the dose data from 32 hospitals were analysed. The third quartile entrance surface dose, dose area product (DAP), weighted CT dose index (CTDIw) and mean glandular dose (MCD) were reported. All the estimated doses were less than the stated International Atomic Energy Agency (IAEA) reference levels for radiographic examinations. However, DAPs for the fluoroscopic examinations had higher dose values than the IAEA reference levels. In addition, the CTDIw and MGD were lower than the IAEA reference levels. [ABSTRACT FROM AUTHOR]
- Subjects :
- RADIATION
MEDICAL radiography
PATIENTS
HOSPITALS
Subjects
Details
- Language :
- English
- ISSN :
- 01448420
- Volume :
- 123
- Issue :
- 4
- Database :
- Complementary Index
- Journal :
- Radiation Protection Dosimetry
- Publication Type :
- Academic Journal
- Accession number :
- 25575808
- Full Text :
- https://doi.org/10.1093/rpd/ncl501