Back to Search Start Over

Development and Validation of HPTLC Method for the Estimation of Etoricoxib.

Authors :
Shah, N. J.
Shah, S. J.
Patel, D. M.
Patel, N. M.
Source :
Indian Journal of Pharmaceutical Sciences; 2006, Vol. 68 Issue 6, p788-789, 2p
Publication Year :
2006

Abstract

A simple, precise, accurate and rapid high performance thin layer chromatographic method has been developed and validated for the determination of etoricoxib in dosage forms. The stationary phase used was precoated silica gel 60F<subscript>254</subscript>. The mobile phase used was a mixture of chloroform: methanol: toluene (4:2:4 v/v). The detection of spot was carried out at 289 nm. The calibration curve was found to be linear between 100 to 600 ng/spot for etoricoxib. The proposed method can be used to determine the drug content of marketed formulations. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0250474X
Volume :
68
Issue :
6
Database :
Complementary Index
Journal :
Indian Journal of Pharmaceutical Sciences
Publication Type :
Academic Journal
Accession number :
24496641
Full Text :
https://doi.org/10.4103/0250-474X.31016