Back to Search Start Over

A NEW HUMAN EYE MODEL FOR OPHTHALMIC BRACHYTHERAPY DOSIMETRY.

Authors :
Yoriyaz, Hélio
Sanchez, Andrea
dos Santos, Adimir
Source :
Radiation Protection Dosimetry; 2005, Vol. 115 Issue 1-4, p316-319, 4p
Publication Year :
2005

Abstract

The present work proposes a new mathematical eye model for ophthalmic brachytherapy dosimetry. This new model includes detailed description of internal structures that were not treated in previous works, allowing dose determination in different regions of the eye for a more adequate clinical analysis. Dose calculations were determined with the MCNP-4C Monte Carlo particle transport code running n parallel environment using PVM. The Amersham CKA4 ophthalmic applicator has been chosen and the depth dose distribution has been determined and compared to those provide by the manufacturer. The results have shown excellent agreement. Besides, absorbed dose values due to both <superscript>125</superscript>I seeds and <superscript>60</superscript>Co plaques were obtained for each one of the different structures which compose the eye model and can give relevant information in eventual clinical analyses. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01448420
Volume :
115
Issue :
1-4
Database :
Complementary Index
Journal :
Radiation Protection Dosimetry
Publication Type :
Academic Journal
Accession number :
20074194
Full Text :
https://doi.org/10.1093/rpd/nci105