Back to Search Start Over

Error Analysis in Two-Terminal Impedance Measurements With Residual Correction.

Authors :
Torrents, Josep M.
Paliàs-Areny, Ramon
Source :
IEEE Transactions on Instrumentation & Measurement; Oct2005, Vol. 54 Issue 5, p2113-2116, 4p
Publication Year :
2005

Abstract

Residual impedance correction in impedance analyzers when using an asymmetrical test fixture needs three reference measurements, usually open circuit, short circuit, and load (meaning an impedance close to the impedance under test). This paper provides an error estimate for impedance measurements that apply a simple open/short correction in spite of using an asymmetrical test fixture. Experimental results show that the minimal error is obtained for impedance values close to the geometric mean of the short-circuit and open-circuit impedances, and that the theoretical prediction is indeed an upper limit for the actual error. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189456
Volume :
54
Issue :
5
Database :
Complementary Index
Journal :
IEEE Transactions on Instrumentation & Measurement
Publication Type :
Academic Journal
Accession number :
18592757
Full Text :