Investigation of Radiation Exposure of Medical Staff During Lateral Fluoroscopy for Posterior Spinal Fusion Surgery.
- Source :
- Journal of Clinical Medicine; Nov2024, Vol. 13 Issue 21, p6442, 10p
- Publication Year :
- 2024
-
Abstract
- Background/Objectives: In spinal surgery, it is especially crucial to insert implants in the correct location. Intraoperative fluoroscopy is often necessary to safely perform spinal surgery because of serious complications that can occur if the screw deviates. However, the use of intraoperative fluoroscopy comes at the cost of radiation exposure to the surgeons and operating room staff. Therefore, it is desirable for spinal surgeons to understand the characteristics of radiation in order to minimize patient and medical staff exposure. This study aimed to create an aerial radiation dose distribution map for lateral fluoroscopy, a commonly used technique for posterior spinal fusion. Methods: A human body-equivalent phantom was placed in a prone position on the Jackson Table. The measurement method used was a lateral fluoroscopic evaluation, assuming posterior spinal fusion. Measurements were taken at three levels: 80 (gonadal), 100 (thoracoabdominal), and 150 cm (lens and thyroid). Results: The highest radiation doses were received by primary surgeons. The scrub nurse was the next most exposed. Conclusions: We developed an aerial dose distribution map for lateral fluoroscopy in posterior spinal fusion. Radiation exposure was the highest among primary surgeons. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 20770383
- Volume :
- 13
- Issue :
- 21
- Database :
- Complementary Index
- Journal :
- Journal of Clinical Medicine
- Publication Type :
- Academic Journal
- Accession number :
- 180783246
- Full Text :
- https://doi.org/10.3390/jcm13216442