Back to Search
Start Over
Offlabel use of Medtronic MiniMed 780G in the management of cystic fibrosis related diabetes in people requiring insulin total daily doses below 8 units: encouraging data from our population.
- Source :
- Acta Diabetologica; Nov2024, Vol. 61 Issue 11, p1483-1489, 7p
- Publication Year :
- 2024
-
Abstract
- Cystic fibrosis (CF)-related diabetes (CFRD), characterized by partial to complete impaired insulin secretion, is the most common extra-pulmonary complication of CF. Actually, insulin is the only approved therapy for its management. Advanced hybrid closed loop (AHCL) systems are the gold standard therapy for type 1 diabetes and have been proposed for other insulin-dependent forms of diabetes, including CFRD. With AHCL systems, people with CFRD can better manage several typical disease-related issues, such as minimal insulin requirements, its variability due to exacerbations or concomitant steroid therapies, nutritional behaviors, the co-existence of CF complications as intestinal malabsorption or liver disease. SmartGuard, the AHCL system for Medtronic Minimed 780G, requires a minimum of 8 units per day to operate. In this paper, we expose a case of two young women with CFRD with total daily insulin requirements < 8 UI, using off-label SmartGuard system over a 3 years of follow-up period, suggesting an evaluation of its use also in people with minimal insulin needs, considering its beneficial impact in glucose control and quality of life. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 09405429
- Volume :
- 61
- Issue :
- 11
- Database :
- Complementary Index
- Journal :
- Acta Diabetologica
- Publication Type :
- Academic Journal
- Accession number :
- 180626490
- Full Text :
- https://doi.org/10.1007/s00592-024-02329-7