Back to Search
Start Over
Regarding Bioanalysis Lasting a Few Minutes: Automated Cooling-SPME and Fast-GC for Urinary 2-Phenyl-2-Propanol Monitoring.
- Source :
- Toxics; Oct2024, Vol. 12 Issue 10, p743, 14p
- Publication Year :
- 2024
-
Abstract
- An innovative SPME head space GC–MS method, in cooling mode, using a fully automated routine, was developed to detect 2-phenyl-2-propanol, a representative urinary metabolite of cumene. Following an acid hydrolysis and derivatization step with lowered quantities of reagents, acetic anhydride and pyridine, a 30 μm polydimethylsiloxane SPME fiber was used to sample derivatized 2-phenyl-2-propanol, such as benzenemethanol,α,α-dimethyl-acetate, from the headspace. Performances of the method, optimized through experimental design, provide an LOD of 0.034 mg/L and an LOQ 0.10 mg/L, with a short sampling time necessary per sample. The method, developed on standard solutions, will be applied to both occupationally exposed and non-exposed populations. [ABSTRACT FROM AUTHOR]
- Subjects :
- ACETIC anhydride
DERIVATIZATION
CUMENE
EXPERIMENTAL design
POLYDIMETHYLSILOXANE
Subjects
Details
- Language :
- English
- ISSN :
- 23056304
- Volume :
- 12
- Issue :
- 10
- Database :
- Complementary Index
- Journal :
- Toxics
- Publication Type :
- Academic Journal
- Accession number :
- 180529874
- Full Text :
- https://doi.org/10.3390/toxics12100743