Back to Search Start Over

The development of 100 Ω cryogenic capsule platinum resistance thermometers for 13 K to 533 K.

Authors :
Xing, Yu Raynee
Xing, Chaoying
Source :
AIP Conference Proceedings; 2024, Vol. 3230 Issue 1, p1-5, 5p
Publication Year :
2024

Abstract

Advanced Sensing Products (ASP) has developed two new miniature capsule platinum resistance thermometers (PRT) for cryogenic use. After testing of a 25.5 Ω prototype of the SSP200 yielded promising results, ASP has developed a specially designed 100 Ω element using a custom diameter platinum wire to manufacture the 100 Ω SSP200 and the PRP200. The SSP200 can be used from a range of 13 K (-260 °C) to 533 K (260 °C). Testing of the 100 Ω SSP200 shows a shift between -2.09 mK to 1.86 mK through 9 cycles at 77 K (liquid nitrogen) and -1.4 mK to 1.4 mK at 273.16 K (triple point of water). Testing performed by Lake Shore Cryotronics, Inc. has shown excellent stability in the 100 Ω SSP200 from 13K to 330K. The SSP200 does not off-gas when tested in ultra-high vacuum at mid 10<superscript>−8</superscript> Torr and is suitable for cryogenic use in vacuum. The PRP200 can be used from a range of 77 K to 477 K. The PRP200 shows a shift between -3.95 mK to 2.56 mK through 9 cycles at 77 K and 0.47 mK to 2.33 mK at 273.16 mK (triple point of water). [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0094243X
Volume :
3230
Issue :
1
Database :
Complementary Index
Journal :
AIP Conference Proceedings
Publication Type :
Conference
Accession number :
180432344
Full Text :
https://doi.org/10.1063/5.0234859