Microwave ablation on ex vivo porcine pancreas: The influence of ablation parameters on ablation results and fat liquefaction.

Authors :
Zhang, Yubo
Gao, Hongjian
He, Huijing
Liu, Ju
Xu, Fan
Wu, Shuicai
Zhou, Jie
Cheng, Zhigang
Source :
Proceedings of the Institution of Mechanical Engineers -- Part H -- Journal of Engineering in Medicine (Sage Publications, Ltd.); Jul2024, Vol. 238 Issue 7, p814-826, 13p
Publication Year :
2024

Abstract

The pancreas is adjacent to critical organs; excessive microwave ablation (MWA) can result in serious complications. The purpose of this paper is to provide the reference data of pancreas MWA for clinicians, analyze the ablation outcomes under different ablation parameters, and determine the critical temperature of pancreatic surface fat liquefaction outflow. Combinations of two power levels (30 W and 55 W), three antenna diameters (1.3 mm, 1.6 mm, and 1.9 mm), and three ablation times (1 min, 1.5 min, and 2 min) were applied to an ex vivo pig pancreas. Temperature measurements were taken at four thermocouple points. The center point is located 5 mm horizontally from the antenna slot, with a temperature measurement point located 5 mm above, below, and to the right of the center point. Main effect analysis and variance analysis were used to quantify the influences of each factor on the ablation outcomes. At 30 W, the antenna diameter contributing the most at 48.5%. At 30 W–1.3 mm–1 min, the spherical index (1.41) is closest to 1. At 55 W, the coagulation zone size was almost only affected by the ablation time, with a contribution rate of 28.7%, the temperature at point C exceeds point B. On the surface of the ex vivo porcine pancreas, the fat outflow temperature was 54ã. Ablation combinations with low power, short duration, and small antenna diameter results in a more nearly spherical coagulation zone. When performing MWA on the pancreas, it is advisable to avoid areas with higher fat content, while keeping the pancreatic surface temperature below 54°C. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09544119
Volume :
238
Issue :
7
Database :
Complementary Index
Journal :
Proceedings of the Institution of Mechanical Engineers -- Part H -- Journal of Engineering in Medicine (Sage Publications, Ltd.)
Publication Type :
Academic Journal
Accession number :
179066589
Full Text :
https://doi.org/10.1177/09544119241261891