Back to Search Start Over

Comparisons between In-Check DIAL® and PF810® in evaluation and training inspiratory capacity for using dry powder inhalers.

Authors :
Wu, Yixing
Li, Li
Gong, Ying
Li, Xingjing
Ye, Xiaofen
Zhang, Jing
Source :
BMC Pulmonary Medicine; 8/2/2024, Vol. 24 Issue 1, p1-11, 11p
Publication Year :
2024

Abstract

Background: Dry powder inhalers (DPIs) rely on both internal resistance and patients' inspiratory capacity for effective operation. Optimal inspiratory technique is crucial for DPI users. This study assessed the accuracy and repeatability of two available devices, PF810® and In-Check DIAL®, and analyzed their measurement errors and consistency in detecting inspiratory capacity. Methods: The accuracy and repeatability of peak inspiratory flow (PIF) and forced inspiratory vital capacity (FIVC) against various internal resistances of the two devices were assessed using standard waveforms generated by a breathing simulator. The agreement of PIF measurements between the two devices in healthy volunteers and chronic obstructive pulmonary disease (COPD) patients was analyzed with the intraclass correlation coefficient and Bland–Altman graphical analysis. Results: PF810® showed great accuracy and repeatability in measuring PIF, except for square waveforms at the lowest flow rate (20 L/min). In-Check DIAL® exhibited poor accuracy against high resistance levels. In scenarios with no resistance, In-Check DIAL® had significantly smaller measurement errors than PF810®, but larger errors against high resistance levels. The two devices showed excellent agreement (ICC > 0.80, P < 0.05), except for healthy volunteers against medium to high resistance (R3-R5) where the ICC was insignificant. Bland–Altman plots indicated small disagreements between the two devices for both healthy volunteers and COPD patients. Conclusions: In-Check DIAL® exhibited poor accuracy and larger measurement errors than PF810® when detecting PIFs against higher internal resistances. However, its good performance against lower internal resistances, along with its cost-effectiveness and convenience made it appropriate for primary care. PF810® showed good accuracy and repeatability and could detect additional parameters of inspiratory capacity beyond PIF, though required further studies to confirm its clinical benefits. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
14712466
Volume :
24
Issue :
1
Database :
Complementary Index
Journal :
BMC Pulmonary Medicine
Publication Type :
Academic Journal
Accession number :
178807078
Full Text :
https://doi.org/10.1186/s12890-024-03191-7