Back to Search Start Over

NUMERICAL CALCULATION AND ANALYSIS OF TEMPERATURE FIELD IN ULTRASONIC WELDING OF POLYPROPYLENE DIALYZER.

Authors :
Song LIU
Ling PENG
Xian HUANG
Bing Rong LIU
Yan GAO
Yi Han WANG
Source :
Thermal Science; 2024, Vol. 28 Issue 3B, p2447-2457, 11p
Publication Year :
2024

Abstract

Ultrasonic plastic welding is widely used in the bonding process of medical device accessories. In this paper, a thermo-force indirect coupling finite element analysis model was established in the ultrasonic plastic welding process between the blood cap and the shell of polypropylene (PP) dialyzer. The temperature field distribution between the blood cap and the shell was simulated and analyzed by using finite element analysis software, and the influence of welding process parameters on the temperature field was studied. The results show that: by changing the ultrasonic amplitude parameters, welding time parameters, initial pressure, etc., the longer the ultrasonic welding time, the temperature of the welding area will increase. In order to ensure the quality of the dialyzer, it should be controlled within 0.8-1 seconds. The increase of ultrasonic amplitude will make the welding temperature continue to rise, and in order to avoid poor welding, the amplitude should not exceed 120 μm. The initial pressure has little effect on the temperature field. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
03549836
Volume :
28
Issue :
3B
Database :
Complementary Index
Journal :
Thermal Science
Publication Type :
Academic Journal
Accession number :
178609195
Full Text :
https://doi.org/10.2298/TSCI231213047L