Bias‐reduced neural networks for parameter estimation in quantitative MRI.

Authors :
Mao, Andrew
Flassbeck, Sebastian
Assländer, Jakob
Source :
Magnetic Resonance in Medicine; Oct2024, Vol. 92 Issue 4, p1638-1648, 11p
Publication Year :
2024

Abstract

Purpose: To develop neural network (NN)‐based quantitative MRI parameter estimators with minimal bias and a variance close to the Cramér–Rao bound. Theory and Methods: We generalize the mean squared error loss to control the bias and variance of the NN's estimates, which involves averaging over multiple noise realizations of the same measurements during training. Bias and variance properties of the resulting NNs are studied for two neuroimaging applications. Results: In simulations, the proposed strategy reduces the estimates' bias throughout parameter space and achieves a variance close to the Cramér–Rao bound. In vivo, we observe good concordance between parameter maps estimated with the proposed NNs and traditional estimators, such as nonlinear least‐squares fitting, while state‐of‐the‐art NNs show larger deviations. Conclusion: The proposed NNs have greatly reduced bias compared to those trained using the mean squared error and offer significantly improved computational efficiency over traditional estimators with comparable or better accuracy. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
07403194
Volume :
92
Issue :
4
Database :
Complementary Index
Journal :
Magnetic Resonance in Medicine
Publication Type :
Academic Journal
Accession number :
178558350
Full Text :
https://doi.org/10.1002/mrm.30135