Prompt gamma energy integration: a new method for online-range verification in proton therapy with pulsed-beams.

Authors :
Everaere, Pierre
Dauvergne, Denis
Gallin-Martel, Marie-Laure
Hérault, Joël
Koudia, Ayoub
Koumeir, Charbel
Létang, Jean Michel
Testa, Étienne
Source :
Frontiers in Physics; 2024, p1-12, 12p
Publication Year :
2024

Abstract

Introduction: We propose a method for prompt-gamma verification of proton range during particle therapy, called Prompt-Gamma Energy Integration (PGEI). Method: This method is based on the measurement of the total energy deposited in a set of detectors located around a patient. It is particularly suited in the case of high-instantaneous beam intensities, like for pulsed beams extracted from a synchro-cyclotron. GATE simuations were used to evaluate the sensitivity, and dedicated scintillators were tested as a function of beam intensity. Results and discussion: Simulations show that millimetric range shifts can be measured at a beam-spot scale. The sensitivity is slightly degraded as compared to the Prompt-Gamma Peak Integration Method, for which Time-of-Flight can be employed to reduce the background in single-photon detection conditions at cyclotron accelerators. Experimentally, lead tungstate scintillators have shown to cope with the high instantaneous gamma count rates for PGEI at synchrocyclotrons. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
2296424X
Database :
Complementary Index
Journal :
Frontiers in Physics
Publication Type :
Academic Journal
Accession number :
178112336
Full Text :
https://doi.org/10.3389/fphy.2024.1371015