Assessing the sensitivity and suitability of a range of detectors for SIMT PSQA.

Authors :
Dunn, Leon
Tamborriello, Alessia
Subramanian, Brindha
Xu, Xiaolei
Ruruku, Tyrone Te
Source :
Journal of Applied Clinical Medical Physics; May2024, Vol. 25 Issue 5, p1-21, 21p
Publication Year :
2024

Abstract

Purpose: Single‐isocenter multi‐target intracranial stereotactic radiotherapy (SIMT) is an effective treatment for brain metastases with complex treatment plans and delivery optimization necessitating rigorous quality assurance. This work aims to assess five methods for quality assurance of SIMT treatment plans in terms of their suitability and sensitivity to delivery errors. Methods: Sun Nuclear ArcCHECK and SRS MapCHECK, GafChromic EBT Radiochromic Film, machine log files, and Varian Portal Dosimetry were all used to measure 15 variations of a single SIMT plan. Variations of the original plan were created with Python. They comprised various degrees of systematic MLC offsets per leaf up to 2 mm, random per‐leaf variations with differing minimum and maximum magnitudes, simulated collimator, and dose miscalibrations (MU scaling). The erroneous plans were re‐imported into Eclipse and plan‐quality degradation was assessed by comparing each plan variation to the original clinical plan in terms of the percentage of clinical goals passing relative to the original plan. Each erroneous plan could be then ranked by the plan‐quality degradation percentage following recalculation in the TPS so that the effects of each variation could be correlated with γ pass rates and detector suitability. Results & conclusions: It was found that 2%/1 mm is a good starting point for the ArcCHECK, Portal Dosimetry, and the SRS MapCHECK methods, respectively, and provides clinically relevant error detection sensitivity. Looser dose criteria of 5%/1 mm or 5%/1.5 mm are suitable for film dosimetry and log‐file‐based methods. The statistical methods explored can be expanded to other areas of patient‐specific QA and detector assessment. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
15269914
Volume :
25
Issue :
5
Database :
Complementary Index
Journal :
Journal of Applied Clinical Medical Physics
Publication Type :
Academic Journal
Accession number :
177192787
Full Text :
https://doi.org/10.1002/acm2.14343