Back to Search
Start Over
Study on the hemodynamic effects of different pulsatile working modes of a rotary blood pump using a microfluidic platform that realizes in vitro cell culture effectively.
- Source :
- Lab on a Chip; 5/7/2024, Vol. 24 Issue 9, p2428-2439, 12p
- Publication Year :
- 2024
-
Abstract
- Rotary blood pumps (RBPs) operating at a constant speed generate non-physiologic blood pressure and flow rate, which can cause endothelial dysfunction, leading to adverse clinical events in peripheral blood vessels and other organs. Notably, pulsatile working modes of the RBP can increase vascular pulsatility to improve arterial endothelial function. However, the laws and related mechanisms of differentially regulating arterial endothelial function under different pulsatile working modes are still unclear. This knowledge gap hinders the optimal selection of the RBP working modes. To address these issues, this study developed a multi-element in vitro endothelial cell culture system (ECCS), which could realize in vitro cell culture effectively and accurately reproduce blood pressure, shear stress, and circumferential strain in the arterial endothelial microenvironment. Performance of this proposed ECCS was validated with numerical simulation and flow experiments. Subsequently, this study investigated the effects of four different pulsation frequency modes that change once every 1-4-fold cardiac cycles (80, 40, 80/3, and 20 cycles per min, respectively) of the RBP on the expression of nitric oxide (NO) and reactive oxygen species (ROS) in endothelial cells. Results indicated that the 2-fold and 3-fold cardiac cycles significantly increased the production of NO and prevented the excessive generation of ROS, potentially minimizing the occurrence of endothelial dysfunction and related adverse events during the RBP support, and were consistent with animal study findings. In general, this study may provide a scientific basis for the optimal selection of the RBP working modes and potential treatment options for heart failure. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 14730197
- Volume :
- 24
- Issue :
- 9
- Database :
- Complementary Index
- Journal :
- Lab on a Chip
- Publication Type :
- Academic Journal
- Accession number :
- 177132423
- Full Text :
- https://doi.org/10.1039/d4lc00159a