Back to Search Start Over

Monitoring of Hip Joint Forces and Physical Activity after Total Hip Replacement by an Integrated Piezoelectric Element.

Authors :
Geiger, Franziska
Bathel, Henning
Spors, Sascha
Bader, Rainer
Kluess, Daniel
Source :
Technologies (2227-7080); Apr2024, Vol. 12 Issue 4, p51, 14p
Publication Year :
2024

Abstract

Resultant hip joint forces can currently only be recorded in situ in a laboratory setting using instrumented total hip replacements (THRs) equipped with strain gauges. However, permanent recording is important for monitoring the structural condition of the implant, for therapeutic purposes, for self-reflection, and for research into managing the predicted increasing number of THRs worldwide. Therefore, this study aims to investigate whether a recently proposed THR with an integrated piezoelectric element represents a new possibility for the permanent recording of hip joint forces and the physical activities of the patient. Hip joint forces from nine different daily activities were obtained from the OrthoLoad database and applied to a total hip stem equipped with a piezoelectric element using a uniaxial testing machine. The forces acting on the piezoelectric element were calculated from the generated voltages. The correlation between the calculated forces on the piezoelectric element and the applied forces was investigated, and the regression equations were determined. In addition, the voltage outputs were used to predict the activity with a random forest classifier. The coefficient of determination between the applied maximum forces on the implant and the calculated maximum forces on the piezoelectric element was R<superscript>2</superscript> = 0.97 (p < 0.01). The maximum forces on the THR could be determined via activity-independent determinations with a deviation of 2.49 ± 13.16% and activity-dependent calculation with 0.87 ± 7.28% deviation. The activities could be correctly predicted using the classification model with 95% accuracy. Hence, piezoelectric elements integrated into a total hip stem represent a promising sensor option for the energy-autonomous detection of joint forces and physical activities. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
22277080
Volume :
12
Issue :
4
Database :
Complementary Index
Journal :
Technologies (2227-7080)
Publication Type :
Academic Journal
Accession number :
176875763
Full Text :
https://doi.org/10.3390/technologies12040051