Complete Prevention of Bubbles in a PDMS-Based Digital PCR Chip with a Multifunction Cavity.

Authors :
Gao, Shiyuan
Xu, Tiegang
Wu, Lei
Zhu, Xiaoyue
Wang, Xuefeng
Chen, Ying
Li, Gang
Li, Xinxin
Source :
Biosensors (2079-6374); Mar2024, Vol. 14 Issue 3, p114, 14p
Publication Year :
2024

Abstract

In a chamber-based digital PCR (dPCR) chip fabricated with polydimethylsiloxane (PDMS), bubble generation in the chambers at high temperatures is a critical issue. Here, we found that the main reason for bubble formation in PDMS chips is the too-high saturated vapor pressure of water at an elevated temperature. The bubbles should be completely prevented by reducing the initial pressure of the system to under 13.6 kPa to eliminate the effects of increased-pressure water vapor. Then, a cavity was designed and fabricated above the PCR reaction layer, and Parylene C was used as a shell covering the chip. The cavity was used for the negative generator in sample loading, PDMS degassing, PCR solution degassing in the digitization process and water storage in the thermal reaction process. The analysis was confirmed and finally achieved a desirable bubble-free, fast-digitization, valve-free and no-tubing connection dPCR. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20796374
Volume :
14
Issue :
3
Database :
Complementary Index
Journal :
Biosensors (2079-6374)
Publication Type :
Academic Journal
Accession number :
176273072
Full Text :
https://doi.org/10.3390/bios14030114