Reduction of cone‐beam CT artifacts in a robotic CBCT device using saddle trajectories with integrated infrared tracking.

Authors :
Wei, Chengtao
Albrecht, Johanna
Rit, Simon
Laurendeau, Matthieu
Thummerer, Adrian
Corradini, Stefanie
Belka, Claus
Steininger, Philipp
Ginzinger, Felix
Kurz, Christopher
Riboldi, Marco
Landry, Guillaume
Source :
Medical Physics; Mar2024, Vol. 51 Issue 3, p1674-1686, 13p
Publication Year :
2024

Abstract

Background: Cone beam computed tomography (CBCT) is widely used in many medical fields. However, conventional CBCT circular scans suffer from cone beam (CB) artifacts that limit the quality and reliability of the reconstructed images due to incomplete data. Purpose: Saddle trajectories in theory might be able to improve the CBCT image quality by providing a larger region with complete data. Therefore, we investigated the feasibility and performance of saddle trajectory CBCT scans and compared them to circular trajectory scans. Methods: We performed circular and saddle trajectory scans using a novel robotic CBCT scanner (Mobile ImagingRing (IRm); medPhoton, Salzburg, Austria). For the saddle trajectory, the gantry executed yaw motion up to ±10∘$\pm 10^{\circ }$ using motorized wheels driving on the floor. An infrared (IR) tracking device with reflective markers was used for online geometric calibration correction (mainly floor unevenness). All images were reconstructed using penalized least‐squares minimization with the conjugate gradient algorithm from RTK with 0.5×0.5×0.5mm3$0.5 \times 0.5\times 0.5 \text{ mm}^3$ voxel size. A disk phantom and an Alderson phantom were scanned to assess the image quality. Results were correlated with the local incompleteness value represented by tan(ψ)$\tan (\psi)$, which was calculated at each voxel as a function of the source trajectory and the voxel's 3D coordinates. We assessed the magnitude of CB artifacts using the full width half maximum (FWHM) of each disk profile in the axial center of the reconstructed images. Spatial resolution was also quantified by the modulation transfer function at 10% (MTF10). Results: When using the saddle trajectory, the region without CB artifacts was increased from 43 to 190 mm in the SI direction compared to the circular trajectory. This region coincided with low values for tan(ψ)$\tan (\psi)$. When tan(ψ)$\tan (\psi)$ was larger than 0.02, we found there was a linear relationship between the FWHM and tan(ψ)$\tan (\psi)$. For the saddle, IR tracking allowed the increase of MTF10 from 0.37 to 0.98 lp/mm. Conclusions: We achieved saddle trajectory CBCT scans with a novel CBCT system combined with IR tracking. The results show that the saddle trajectory provides a larger region with reliable reconstruction compared to the circular trajectory. The proposed method can be used to evaluate other non‐circular trajectories. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00942405
Volume :
51
Issue :
3
Database :
Complementary Index
Journal :
Medical Physics
Publication Type :
Academic Journal
Accession number :
175989222
Full Text :
https://doi.org/10.1002/mp.16943