A discussion of "In vitro analysis of shear stress: CAD milled vs printed denture base resins with bonded denture tooth".

Authors :
Righesso, Leonardo Augusto Rachele
Smith, Clark
Source :
Journal of Prosthodontics; Mar2024, Vol. 33 Issue 3, p203-205, 3p
Publication Year :
2024

Abstract

The article discusses a study comparing the shear stress of different denture base resins with bonded denture teeth. The authors express concern about the methodology used in the study, specifically regarding the bonding of teeth to the denture bases. They argue that the failure to follow the validated workflow outlined in the Instructions for Use (IFU) impacted the results. The authors highlight conflicting results from previous studies and emphasize the importance of following the approved IFU for optimal results. They conclude that the protocol described in the study should not be used for patients and recommend following the validated workflow for a stronger bond and high clinical performance. [Extracted from the article]