Comparative study on measurements of radiochromic films using portable colorimeters.
- Source :
- Scientific Reports; 2/9/2024, Vol. 14 Issue 1, p1-9, 9p
- Publication Year :
- 2024
-
Abstract
- The recently proposed method for on-site radiation dosimetry by a combination of radiochromic film and portable colorimeter was tested for six combinations of two popular Gafchromic films (EBT3 and EBT-XD) and three commercially available portable colorimeters (nix pro2, nix spectro2 and Spectro1 Pro; abbreviated to "NixP", "NixS" and "SpoP", respectively). EBT3 and EBT-XD were irradiated with X-rays (160 kV, 6.3 mA) up to 40 Gy and 80 Gy, respectively, and the radiation-induced color levels of RGB and CMYK components were measured with the three colorimeters. Angle dependence was examined by reading at 15° intervals. As a result, it was judged that all combinations would work effectively under certain irradiation conditions. NixP and NixS were applicable to a wider dose range for both films, while SpoP fit a lower dose range. On the other hand, SpoP showed an advantageous feature of no angular dependence in reading films, while NixP and NixS showed significant angle-dependent changes. These differences are considered to be attributable to the different geometries of LED light emission, which came from all directions (360°) in SpoP, 4 directions in NixP, and 8 directions in NixS. These findings are expected to expand the applicability of the novel method to various occasions of on-site dosimetry. [ABSTRACT FROM AUTHOR]
- Subjects :
- COLORIMETERS
RADIATION dosimetry
THERMOLUMINESCENCE dosimetry
COMPARATIVE studies
Subjects
Details
- Language :
- English
- ISSN :
- 20452322
- Volume :
- 14
- Issue :
- 1
- Database :
- Complementary Index
- Journal :
- Scientific Reports
- Publication Type :
- Academic Journal
- Accession number :
- 175831929
- Full Text :
- https://doi.org/10.1038/s41598-024-54017-9