Back to Search
Start Over
A silicon photomultiplier based compact gamma spectrometer for environmental gamma radiation monitoring networks.
- Source :
- Radiation Protection Dosimetry; Jan2024, Vol. 200 Issue 1, p1-11, 11p
- Publication Year :
- 2024
-
Abstract
- In this paper, the details of the development and performance characterisation of a compact, low-power gamma spectrometer for environmental gamma radiation monitoring networks are presented. To reduce the power consumption and the size of the spectrometer, a gamma detector comprising a silicon photomultiplier coupled to a Gd<subscript>3</subscript>Ga<subscript>3</subscript>Al<subscript>2</subscript>O<subscript>12</subscript>:Ce,B (GGAG:Ce,B) scintillator has been used for gamma spectrometry. Initially, a Monte Carlo simulation study was carried out to verify the suitability of the 5 mm × 5 mm × 5 mm GGAG:Ce,B crystal for spectrometry of gamma sources in the energy range 60–1332 keV. For minimising the power consumption, the signal processing electronics has been custom designed. This electronics was realised using standard off-the-shelf components to reduce the cost. The developed spectrometer is of size 16 cm × 10 cm × 6 cm, weighs 600 g and consumes 600 mW power. The spectrometer is developed such that it could be directly interfaced with GSM/Xbee for wireless communication with the radiation monitoring networks. The lower-level discriminator threshold of the system is 40 keV and the total electronic noise is <20 keV. The experimentally measured sensitivity of the spectrometer for <superscript>137</superscript>Cs (662 keV) is 2.4 cps/μGy/h at 3.5 V overvoltage. The spectrometer offers excellent linearity over the measured energy range of 60–1332 keV and an energy resolution of ~10% for 662 keV gamma-ray at room temperature. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 01448420
- Volume :
- 200
- Issue :
- 1
- Database :
- Complementary Index
- Journal :
- Radiation Protection Dosimetry
- Publication Type :
- Academic Journal
- Accession number :
- 174559646
- Full Text :
- https://doi.org/10.1093/rpd/ncad256