Analysis of copper metal with a K-Alpha instrument from Thermo Scientific by XPS at 60 and 200 eV pass energy.

Authors :
Dean, Annika M.
Jafari, Samira
Linford, Matthew R.
Source :
Surface Science Spectra; Dec2023, Vol. 30 Issue 2, p1-11, 11p
Publication Year :
2023

Abstract

Copper metal was analyzed by x-ray photoelectron spectroscopy (XPS) using a K-Alpha instrument from Thermo Scientific that employed an Al K-alpha x-ray source (1486.6 eV). The sample was the copper standard on the instrument sample stage that was sputtered with Ar<superscript>+</superscript> to remove its oxide layer and hydrocarbon contamination. Shown in this work are the survey spectrum and high-resolution Cu 2p, Cu L<subscript>3</subscript>M<subscript>45</subscript>M<subscript>45</subscript>, Cu 3s, Cu 3p, O 1s, C 1s, and Ar 2p narrow scans. Narrow and survey scans were collected at 60 and 200 eV pass energy, respectively. [ABSTRACT FROM AUTHOR]