Evaluation of a Balloon Implant for Simultaneous Magnetic Nanoparticle Hyperthermia and High-Dose-Rate Brachytherapy of Brain Tumor Resection Cavities.

Authors :
Wan, Shuying
Rodrigues, Dario B.
Kwiatkowski, Janet
Khanna, Omaditya
Judy, Kevin D.
Goldstein, Robert C.
Overbeek Bloem, Marty
Yu, Yan
Rooks, Sophia E.
Shi, Wenyin
Hurwitz, Mark D.
Stauffer, Paul R.
Source :
Cancers; Dec2023, Vol. 15 Issue 23, p5683, 19p
Publication Year :
2023

Abstract

Simple Summary: Glioblastoma (GBM) generally recurs locally with a dismal median survival of <18 months. Combining thermal therapy with radiation therapy enhances radiation response and improves clinical outcomes. This study evaluates a thermobrachytherapy balloon implant intended for the simultaneous heat and radiation treatment of tumor resection cavities. The data demonstrate that our prototype implant produces spherically symmetric heat and radiation dose distributions around the balloon, while the in vivo experiments confirm our ability to heat ≥40 °C at a 5 mm distance from the balloon surface in highly perfused pig brain tissue. The device is now ready for the finalization of regulatory approvals in anticipation of early-stage clinical investigation. Previous work has reported the design of a novel thermobrachytherapy (TBT) balloon implant to deliver magnetic nanoparticle (MNP) hyperthermia and high-dose-rate (HDR) brachytherapy simultaneously after brain tumor resection, thereby maximizing their synergistic effect. This paper presents an evaluation of the robustness of the balloon device, compatibility of its heat and radiation delivery components, as well as thermal and radiation dosimetry of the TBT balloon. TBT balloon devices with 1 and 3 cm diameter were evaluated when placed in an external magnetic field with a maximal strength of 8.1 kA/m at 133 kHz. The MNP solution (nanofluid) in the balloon absorbs energy, thereby generating heat, while an HDR source travels to the center of the balloon via a catheter to deliver the radiation dose. A 3D-printed human skull model was filled with brain-tissue-equivalent gel for in-phantom heating and radiation measurements around four 3 cm balloons. For the in vivo experiments, a 1 cm diameter balloon was surgically implanted in the brains of three living pigs (40–50 kg). The durability and robustness of TBT balloon implants, as well as the compatibility of their heat and radiation delivery components, were demonstrated in laboratory studies. The presence of the nanofluid, magnetic field, and heating up to 77 °C did not affect the radiation dose significantly. Thermal mapping and 2D infrared images demonstrated spherically symmetric heating in phantom as well as in brain tissue. In vivo pig experiments showed the ability to heat well-perfused brain tissue to hyperthermic levels (≥40 °C) at a 5 mm distance from the 60 °C balloon surface. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20726694
Volume :
15
Issue :
23
Database :
Complementary Index
Journal :
Cancers
Publication Type :
Academic Journal
Accession number :
174115432
Full Text :
https://doi.org/10.3390/cancers15235683