Eddy current‐induced artifact correction in high b‐value ex vivo human brain diffusion MRI with dynamic field monitoring.

Authors :
Ramos‐Llordén, Gabriel
Park, Daniel J.
Kirsch, John E.
Scholz, Alina
Keil, Boris
Maffei, Chiara
Lee, Hong‐Hsi
Bilgic, Berkin
Edlow, Brian L.
Mekkaoui, Choukri
Yendiki, Anastasia
Witzel, Thomas
Huang, Susie Y.
Source :
Magnetic Resonance in Medicine; Feb2024, Vol. 91 Issue 2, p541-557, 17p
Publication Year :
2024

Abstract

Purpose: To investigate whether spatiotemporal magnetic field monitoring can correct pronounced eddy current‐induced artifacts incurred by strong diffusion‐sensitizing gradients up to 300 mT/m used in high b‐value diffusion‐weighted (DW) EPI. Methods: A dynamic field camera equipped with 16 1H NMR field probes was first used to characterize field perturbations caused by residual eddy currents from diffusion gradients waveforms in a 3D multi‐shot EPI sequence on a 3T Connectom scanner for different gradient strengths (up to 300 mT/m), diffusion directions, and shots. The efficacy of dynamic field monitoring‐based image reconstruction was demonstrated on high‐gradient strength, submillimeter resolution whole‐brain ex vivo diffusion MRI. A 3D multi‐shot image reconstruction framework was developed that incorporated the nonlinear phase evolution measured with the dynamic field camera. Results: Phase perturbations in the readout induced by residual eddy currents from strong diffusion gradients are highly nonlinear in space and time, vary among diffusion directions, and interfere significantly with the image encoding gradients, changing the k‐space trajectory. During the readout, phase modulations between odd and even EPI echoes become non‐static and diffusion encoding direction‐dependent. Superior reduction of ghosting and geometric distortion was achieved with dynamic field monitoring compared to ghosting reduction approaches such as navigator‐ and structured low‐rank‐based methods or MUSE followed by image‐based distortion correction with the FSL tool "eddy." Conclusion: Strong eddy current artifacts characteristic of high‐gradient strength DW‐EPI can be well corrected with dynamic field monitoring‐based image reconstruction. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
07403194
Volume :
91
Issue :
2
Database :
Complementary Index
Journal :
Magnetic Resonance in Medicine
Publication Type :
Academic Journal
Accession number :
173925242
Full Text :
https://doi.org/10.1002/mrm.29873